Effect of dual-source cardiac computed tomography on patient radiation dose in a clinical setting: Comparison to single-source imaging

  • Sandra S. Halliburton, PhD

      Affiliations

    • Cardiovascular Imaging Laboratory, Imaging Institute, Cleveland Clinic, 9500 Euclid Avenue, Cleveland, OH 44195, USA
    • Department of Biomedical Engineering, Lerner Research Institute, Cleveland Clinic, 9500 Euclid Avenue, Cleveland, OH 44195, USA
    • Corresponding Author InformationCorresponding author.
  • ,
  • Srikanth Sola, MD

      Affiliations

    • Cardiovascular Imaging Laboratory, Imaging Institute, Cleveland Clinic, 9500 Euclid Avenue, Cleveland, OH 44195, USA
    • Department of Cardiovascular Medicine, Heart Institute, Cleveland Clinic, 9500 Euclid Avenue, Cleveland, OH 44195, USA
  • ,
  • Stacie A. Kuzmiak, RT

      Affiliations

    • Cardiovascular Imaging Laboratory, Imaging Institute, Cleveland Clinic, 9500 Euclid Avenue, Cleveland, OH 44195, USA
  • ,
  • Nancy A. Obuchowski, PhD

      Affiliations

    • Department of Quantitative Health Sciences, Lerner Research Institute, Cleveland Clinic, 9500 Euclid Avenue, Cleveland, OH 44195, USA
  • ,
  • Milind Desai, MD

      Affiliations

    • Cardiovascular Imaging Laboratory, Imaging Institute, Cleveland Clinic, 9500 Euclid Avenue, Cleveland, OH 44195, USA
    • Department of Cardiovascular Medicine, Heart Institute, Cleveland Clinic, 9500 Euclid Avenue, Cleveland, OH 44195, USA
  • ,
  • Scott D. Flamm, MD

      Affiliations

    • Cardiovascular Imaging Laboratory, Imaging Institute, Cleveland Clinic, 9500 Euclid Avenue, Cleveland, OH 44195, USA
    • Department of Cardiovascular Medicine, Heart Institute, Cleveland Clinic, 9500 Euclid Avenue, Cleveland, OH 44195, USA
  • ,
  • Paul Schoenhagen, MD

      Affiliations

    • Cardiovascular Imaging Laboratory, Imaging Institute, Cleveland Clinic, 9500 Euclid Avenue, Cleveland, OH 44195, USA
    • Department of Cardiovascular Medicine, Heart Institute, Cleveland Clinic, 9500 Euclid Avenue, Cleveland, OH 44195, USA

Received 2 July 2008; accepted 20 September 2008. published online 10 October 2008.

Background

Dual-source computed tomography (DSCT) was introduced with significant hardware and software changes compared with single-source CT (SSCT), resulting in improved temporal resolution (83 ms) and the potential for improved image quality. The effect of these changes on radiation dose requirements for coronary CT angiography in clinical practice has not been investigated.

Objective

We evaluated patient radiation dose and image quality of electrocardiogram (ECG)–gated helical techniques, using DSCT compared with SSCT for clinical imaging of the coronary arteries.

Methods

DSCT data from 160 patients were evaluated; 82 patients (DSCT group 1) were imaged with early software, and 78 patients (DSCT group 2) were imaged with a later software version. Patients imaged with SSCT (n = 124) were the control group. Effective radiation dose values were estimated for all patients. Image noise was measured, and image quality was evaluated on a 5-point scale.

Results

Effective dose values for DSCT group 2 (11.7 ± 4.0 mSv) were not different from those for SSCT group (10.9 ± 2.9 mSv); the highest doses, 13.2 ± 3.2 mSv, were recorded for DSCT group 1 (P < 0.001). A decrease in image noise was observed for DSCT compared with SSCT (P ≤ 0.001) as was an increase in image quality (P < 0.01). With optimized DSCT imaging, lower dose values were associated with (1) shorter scan range, (2) lower maximum tube current, and (3) lower fraction of R-R interval receiving maximum tube current.

Conclusion

ECG-gated helical DSCT can provide images of the coronary arteries with improved image quality and decreased noise without an increase in radiation dose compared with SSCT in clinical patient groups.

Keywords: Computed tomography, Coronary disease, Dual source, Imaging, Radiation dose

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 Conflict of interest: S.S. Halliburton received research funding from Siemens Medical Solutions for an unrelated project. The other authors report no conflicts of interest.

PII: S1934-5925(08)00616-3

doi:10.1016/j.jcct.2008.09.003

Journal of Cardiovascular Computed Tomography
Volume 2, Issue 6 , Pages 392-400, November 2008