Open AccessPublished:February 06, 2023DOI:https://doi.org/10.1016/j.jcct.2023.01.042

      Abstract

      Background

      The Valve-in-Valve International Data (VIVID) registry proposed a simplified classification to assess the risk of coronary obstruction during valve-in-valve transcatheter aortic valve implantation (ViV-TAVI) based on preprocedural multi-detector computed tomography (MDCT). We investigated the validity of the VIVID classification in patients undergoing ViV-TAVI for degenerated bioprostheses.

      Methods

      Patients undergoing ViV-TAVI for degenerated bioprostheses were prospectively included in this study. The risk of coronary obstruction among patients treated with stented valves was retrospectively evaluated based on anatomical assessment on pre-procedural MDCT.

      Results

      Among a total of 137 patients that underwent ViV-TAVI between August 2007 and June 2021, 109 patients had stented, sutureless, or transcatheter degenerated bioprosthesis of which 96 (88%) had adequate MDCT data for risk assessment. High-risk anatomy for coronary obstruction (VIVID type IIB, IIIB, or IIIC) in either the left or right coronary artery was observed in 30 patients (31.3%). Of the 30 patients with high-risk anatomy, coronary protection using wire protection or BASILICA (bioprosthetic or native aortic scallop intentional laceration to prevent iatrogenic coronary artery obstruction) was performed in 3 patients (10.0%). Three patients treated with stentless valves and one patient treated with a stented valve with externally mounted leaflets had coronary obstruction. None of the patients with high risk anatomy according to MDCT had coronary obstruction even without coronary protection.

      Conclusions

      Coronary obstruction occurred in none of the patients classified as high-risk patients according to the VIVID classification despite the absence of coronary protection. Refined tools are required to assess the risk of coronary obstruction.

      Clinical trial registration

      Graphical abstract

      Keywords

      Abbreviations:

      AS (aortic stenosis), BASILICA (bioprosthetic or native aortic scallop intentional laceration to prevent iatrogenic coronary artery obstruction), ViV-TAVI (valve-in-valve transcatheter aortic valve implantation), VIVID (Valve-in-Valve International Data)

      1. Introduction

      Valve-in-valve transcatheter aortic valve implantions (ViV-TAVI) are expanding as the new standard of care for the treatment of degenerated bioprostheses.
      • Dvir D.
      • Webb J.G.
      • Bleiziffer S.
      • et al.
      Transcatheter aortic valve implantation in failed bioprosthetic surgical valves.
      • Tuzcu E.M.
      • Kapadia S.R.
      • Vemulapalli S.
      • et al.
      Transcatheter aortic valve replacement of failed surgically implanted bioprostheses: the STS/ACC registry.
      • Webb J.G.
      • Mack M.J.
      • White J.M.
      • et al.
      Transcatheter aortic valve implantation within degenerated aortic surgical bioprostheses: PARTNER 2 valve-in-valve registry.
      • Webb J.G.
      • Murdoch D.J.
      • Alu M.C.
      • et al.
      3-Year outcomes after valve-in-valve transcatheter aortic valve replacement for degenerated bioprostheses: the PARTNER 2 registry.
      • Deeb G.M.
      • Chetcuti S.J.
      • Reardon M.J.
      • et al.
      1-Year results in patients undergoing transcatheter aortic valve replacement with failed surgical bioprostheses.
      • Bleiziffer S.
      • Simonato M.
      • Webb J.G.
      • et al.
      Long-term outcomes after transcatheter aortic valve implantation in failed bioprosthetic valves.
      • Hahn R.T.
      • Webb J.
      • Pibarot P.
      • et al.
      5-Year follow-up from the PARTNER 2 aortic valve-in-valve registry for degenerated aortic surgical bioprostheses.
      • Carroll J.D.
      • Mack M.J.
      • Vemulapalli S.
      • et al.
      STS-ACC TVT registry of transcatheter aortic valve replacement.
      In the current guidelines for the management of valvular heart disease, ViV-TAVI assumes a class 2a level of evidence B-NR recommendation.
      • Otto C.M.
      • Nishimura R.A.
      • Bonow R.O.
      • et al.
      ACC/AHA guideline for the management of patients with valvular heart disease: a report of the American college of cardiology/American heart association joint committee on clinical practice guidelines.
      ,
      • Vahanian A.
      • Beyersdorf F.
      • Praz F.
      • et al.
      2021 ESC/EACTS Guidelines for the management of valvular heart disease.
      Coronary obstruction is the most feared complication of ViV-TAVI,
      • Dvir D.
      • Leipsic J.
      • Blanke P.
      • et al.
      Coronary obstruction in transcatheter aortic valve-in-valve implantation: preprocedural evaluation, device selection, protection, and treatment.
      and there is a growing demand to establish a standardized risk assessment method for the prevention of coronary obstruction in patients undergoing ViV-TAVI.
      • Okuno T.
      Risk of "future" coronary obstruction: a key factor in patient-tailored lifetime management of aortic stenosis.
      Prior studies have suggested several anatomical risk factors for coronary obstruction during ViV-TAVI and have proposed techniques to prevent this dreadful complication.
      • Ribeiro H.B.
      • Rodes-Cabau J.
      • Blanke P.
      • et al.
      Incidence, predictors, and clinical outcomes of coronary obstruction following transcatheter aortic valve replacement for degenerative bioprosthetic surgical valves: insights from the VIVID registry.
      • Khan J.M.
      • Greenbaum A.B.
      • Babaliaros V.C.
      • et al.
      The BASILICA trial: prospective multicenter investigation of intentional leaflet laceration to prevent TAVR coronary obstruction.
      • Mercanti F.
      • Rosseel L.
      • Neylon A.
      • et al.
      Chimney stenting for coronary occlusion during TAVR: insights from the chimney registry.
      BASILICA (bioprosthetic or native aortic scallop intentional laceration to prevent iatrogenic coronary artery obstruction) is a recently proposed transcatheter procedure which lacerates the aortic valve leaflets before TAVI, allowing them to splay to the sides after transcatheter heart valve deployment and, therefore, maintaining coronary perfusion. Although the feasibility of the technique has been shown in a dedicated study, appropriate patient selection for the technique, including anatomical considerations, remains largely unknown.
      • Khan J.M.
      • Greenbaum A.B.
      • Babaliaros V.C.
      • et al.
      The BASILICA trial: prospective multicenter investigation of intentional leaflet laceration to prevent TAVR coronary obstruction.
      Recently, the Valve-in-Valve International Data (VIVID) registry proposed a simplified classification based on preprocedural multi-detector computed tomography (MDCT) to assess the risk of coronary obstruction during ViV-TAVI in stented bioprostheses and the possible need for BASILICA.
      • Tang G.H.L.
      • Komatsu I.
      • Tzemach L.
      • et al.
      Risk of coronary obstruction and the need to perform BASILICA: the VIVID classification.
      In the present study, we aimed to investigate the validity of the VIVID classification in patients undergoing ViV-TAVI based on data from a prospective TAVI registry.

      2. Methods

      2.1 Study design and population

      All patients undergoing TAVI at Bern University Hospital, Bern, Switzerland, are consecutively recorded in a prospective institutional database as part of the SwissTAVI registry which is mandated by the Swiss health authorities (registered at clinicaltrials.gov with NCT01368250).
      • Stortecky S.
      • Franzone A.
      • Heg D.
      • et al.
      Temporal trends in adoption and outcomes of transcatheter aortic valve implantation: a SwissTAVI Registry analysis.
      The present analysis included consecutive patients that underwent ViV-TAVI between August 2007 and June 2021. Among them, patients with a stented, sutureless, or transcatheter degenerated bioprosthesis and adequate pre-procedural MDCT images were evaluated for the risk of coronary obstruction according to the VIVID classification. The systolic phase of multi-detector computed tomography with the least motion and blooming artifact was selected. Patients implanted with stentless valves were not evaluated for the risk of coronary obstruction as these valves lack a fluoroscopic marker, and the posts are not identified on MDCT. The registry is approved by the Bern cantonal ethics committee, and patients provided written informed consent to participate.

      2.2 Assessment of coronary obstruction risk

      The risk of coronary obstruction was retrospectively evaluated in patients with stented, sutureless, or transcatheter degenerated bioprostheses based on anatomical assessment on pre-procedural MDCT. The pre-procedural MDCT was performed as previously described. Acquired images were transferred to a dedicated workstation (3mensio Structural Heart, 3mensio Medical Imaging BV, Bilthoven, The Netherlands) and evaluated by experienced imaging specialists in a dedicated Corelab.
      • Okuno T.
      • Brugger N.
      • Asami M.
      • et al.
      Clinical impact of mitral calcium volume in patients undergoing transcatheter aortic valve implantation.
      The risk assessment included 1) identification of the basal ring of the bioprosthesis; 2) assessment of the length of the prosthetic posts in relation to the offtake of the coronary arteries and the height of the coronary sinus; 3) measurement of the valve to coronary ostium (VTC) distance and the valve to sinotubular junction (STJ) (VTSTJ) distance. The basal ring of the bioprosthesis was identified on a transverse double oblique plane. In cases in which the coronary arteries took off below the bioprosthetic posts, VTC was measured. The VTC was assessed using a virtual cylinder with the size of the implanted bioprosthesis. A virtual cylinder with the implanted prosthesis-specific diameter was aligned with the posts of the implanted bioprosthesis. Then, the distance between the edge of the cylinder and the coronary ostium was measured. In cases in which the prosthesis posts extended above the STJ, the VTSTJ was also evaluated (Fig. 1).
      • Ribeiro H.B.
      • Rodes-Cabau J.
      • Blanke P.
      • et al.
      Incidence, predictors, and clinical outcomes of coronary obstruction following transcatheter aortic valve replacement for degenerative bioprosthetic surgical valves: insights from the VIVID registry.
      ,
      • Blanke P.
      • Weir-McCall J.R.
      • Achenbach S.
      • et al.
      Computed tomography imaging in the context of transcatheter aortic valve implantation (TAVI)/Transcatheter aortic valve replacement (TAVR): an expert consensus document of the society of cardiovascular computed tomography.
      ,
      • Tarantini G.
      • Dvir D.
      • Tang G.H.L.
      Transcatheter aortic valve implantation in degenerated surgical aortic valves.
      Each coronary artery ostium was evaluated separately.
      Fig. 1
      Fig. 1Assessment for risk of coronary artery obstruction according to VIVID classification. (A) Identifying the three most basal points of the bioprosthesis. (B) Draw a circular region of interest with implanted valve-specified diameter. (C) Measurement of VTC from the virtual circle to the coronary ostium. (D) Measurement of VTSTJ from the virtual circle to the sinotubular junction. LCA = left coronary artery; VIVID = Valve-in-Valve International Data; VTC ​= ​virtual valve to coronary ostium; VTSTJ ​= ​virtual valve to sinotubular junction.

      2.3 VIVID classification

      We classified each coronary artery according to the VIVID classifi-cation as follows;
      Type I: The failed bioprosthetic leaflets extend below the coronary ostia plane.
      Type II: The failed bioprosthetic leaflets extend between the cor-onary ostium and the STJ. Type II anatomy was further classified as type IIA (VTC ≥4 ​mm) or type IIB (VTC <4 ​mm) based on the VTC distance.
      Type III: The failed bioprosthesis leaflets extend above the level of STJ. Type III anatomy was further classified as type IIIA (VTC ≥4 ​mm and VTSTJ ≥3.5 ​mm), type IIIB (VTC <4 ​mm), or type IIIC (VTC ≥4 ​mm but VTSTJ <3.5 ​mm) based on the VTC and VTSTJ distances.
      Types IIB, IIIB, and IIIC are considered high risk anatomies for coronary obstruction and BASILICA is recommended to be considered to mitigate the risk (Fig. 2).
      • Tang G.H.L.
      • Komatsu I.
      • Tzemach L.
      • et al.
      Risk of coronary obstruction and the need to perform BASILICA: the VIVID classification.
      Fig. 2
      Fig. 2Prevalence of patients at risk of coronary obstruction during ViV-TAVI. Pie chart illustrating the prevalence of patients at high and low risk of coronary obstruction (Left), the prevalence of VIVID categories in the left coronary artery (Upper Right), and the right coronary artery (Lower Right). BASILICA ​= ​bioprosthetic or native aortic scallop intentional laceration to prevent iatrogenic coronary artery obstruction; MDCT ​= ​multi-detector computed tomography; ViV-TAVI ​= ​valve-in-valve transcatheter aortic valve implantation; VIVID = Valve-in-Valve International Data; VTC ​= ​virtual valve to coronary ostium; VTSTJ ​= ​virtual valve to sinotubular junction.

      2.4 Data collection and clinical endpoints

      Baseline clinical, procedural, and follow-up data were prospectively recorded in a web-based database, held at the Clinical Trials Unit of the University of Bern, Switzerland. The decision for transcatheter heart valve type was made by the operator. Coronary protection was performed by heart team decision. Clinical follow-up data were obtained by standardized interviews, documentation from referring physicians, and hospital discharge summaries. All adverse events were systematically collected and adjudicated by a dedicated clinical event committee based on the standardized VARC definitions applicable at the time of the procedure.
      • Leon M.B.
      • Piazza N.
      • Nikolsky E.
      • et al.
      Standardized endpoint definitions for transcatheter aortic valve implantation clinical trials: a consensus report from the valve academic research consortium.
      ,
      • Kappetein A.P.
      • Head S.J.
      • Généreux P.
      • et al.
      Updated standardized endpoint definitions for transcatheter aortic valve implantation: the Valve Academic Research Consortium-2 consensus document.

      2.5 Statistical analysis

      Categorical variables are represented as frequencies and percentages. Continuous variables are presented as mean values ​± ​standard deviation (SD). All statistical analyses were performed using R for Windows 4.0.2 (R Foundation for Statistical Computing, Vienna, Austria).

      3. Results

      3.1 Study population and baseline characteristics

      Among 3201 consecutive patients enrolled in the prospective TAVI registry, 137 patients (4.3%) underwent ViV-TAVI between August 2007 and June 2021 (Fig. 3). Most of the bioprostheses were stented valves (69.4%), followed by stentless valves (20.4%), sutureless valves (5.1%), and transcatheter heart valves (5.1%). Overall, 41 patients (29.9%) were female, the mean age of the cohort was 82 ​± ​7 years, body mass index was 25.6 ​± ​4.4 ​kg/m2, and Society of Thoracic Surgeons Predicted Risk of Mortality score was 6.9 ​± ​4.9 (Table 1).
      Fig. 3
      Fig. 3Study flowchart. AS ​= ​aortic stenosis; MDCT ​= ​multi-detector computed tomography; TAVI ​= ​transcatheter aortic valve implantation; VIVID = Valve-in-Valve International Data.
      Table 1Baseline characteristics.
      All patients (N ​= ​137)Patients with assessment of the VIVID classification (N ​= ​96)
      Age, years82 ​± ​783 ​± ​6
      Female, n (%)41 (29.9)331 (34.4)
      Body mass index, kg/m225.6 ​± ​4.425.5 ​± ​4.6
      LVEF, %50.5 ​± ​15.351.1 ​± ​15.4
      STS-PROM, %6.9 ​± ​4.97.3 ​± ​5.3
      Implanted bioprosthetic valve, n (%)
       Stented95 (69.3)84 (85.4)
      Internally mounted leaflets62 (45.3)58 (60.4)
      Externally mounted leaflets31 (22.6)24 (25.0)
      Type unknown2 (1.5)2 (2.1)
       Stentless28 (20.4)
       Sutureless7 (5.1)7 (7.3)
       Transcatheter heart valve7 (5.1)5 (5.2)
      Values are mean ​± ​SD or n (%).
      LVEF ​= ​left ventricular ejection fraction; STS-PROM = Society of Thoracic Surgeons Predicted Risk of Mortality.

      3.2 Procedural characteristics

      Procedural characteristics are summarized in Table 2. The reasons for ViV-TAVI were bioprosthetic valve stenosis in 53.3%, bioprosthetic valve regurgitation in 37.2%, and both a combination of stenosis and regurgitation in 9.5%. Most of the transcatheter heart valves used for ViV-TAVI were self-expanding valves (71.5%), followed by balloon-expandable valves (27.0%), and mechanically-expanding valves (1.5%). Coronary protection was performed in 13 patients (9.5%): 7 (5.1%) underwent BASILICA (5 single BASILICA for left coronary cusp and 2 double BASILICA for left and right coronary cusps), and 6 (4.4%) had wire protection. Coronary obstruction during the index procedure occurred in 4 patients (2.9%): 3 patients had a stentless valve and 1 patient had a stented valve with externally mounted leaflets.
      Table 2Procedural characteristics.
      All patients (N ​= ​137)Patients with assessment of the VIVID classification (N ​= ​96)
      Reason for ViV-TAVI, n (%)
       Bioprosthetic regurgitation51 (37.2)27 (28.1)
       Bioprosthetic stenosis73 (53.3)59 (61.5)
       Bioprosthetic regurgitation and stenosis13 (9.5)10 (10.4)
      Device type, n (%)
       Balloon-expandable37 (27.0)25 (26.0)
       Self-expanding98 (71.5)71 (74.0)
       Mechanically-expanding2 (1.5)0
      Device size, mm25.8 ​± ​2.425.4 ​± ​2.1
      Coronary protection, n (%)13 (9.5)7 (7.3)
       BASILICA, n (%)7 (5.1)5 (5.2)
       Wire protection, n (%)6 (4.4)2 (2.1)
      Coronary obstruction, n (%)4 (2.9)0
      Values are mean ​± ​SD or n (%).
      BASILICA ​= ​bioprosthetic or native aortic scallop intentional laceration to prevent iatrogenic coronary artery obstruction; ViV-TAVI ​= ​valve-in-valve transcatheter aortic valve implantation; VIVID = Valve-in-Valve International Data.

      3.3 VIVID classification

      After excluding 41 patients (28; stentless bioprostheses and 13; stented bioprostheses with inadequate MDCT images to comprehensively evaluate both coronary artery ostia) and 3 right coronary arteries with inadequate MDCT images, a total of 96 patients with 189 coronary arteries [96 left coronary artery (LCA) and 93 right coronary arteries (RCA)] were assessed for the risk of coronary obstruction according to the VIVID classification. MDCT measurements are summarized in Table 3. Each coronary artery was categorized according to the VIVID classification: With respect to the left coronary artery, VIVID type I was found in 21.9%; type IIA in 27.1%; type IIB in 4.2%; type IIIA in 33.3%; type IIIB in 8.3%; and type IIIC 5.2%; with respect to the right coronary artery, VIVID type I was found in 40.9%; type IIA in 24.7%; type IIB in 6.5%; type IIIA in 17.2%; type IIIB in 7.5%, and type IIIC in 3.2% of coronary arteries (Fig. 2). High-risk anatomy for coronary obstruction in either the LCA or the RCA according to the VIVID classification was observed in 30 patients (31.3%) (Fig. 2). In a sensitivity analysis using a cut-off of <2.5 ​mm for VTSTJ as proposed by the VIVID classification, high-risk anatomy in either the LCA or the RCA was observed in 25 patients (26.0%) (Table 3).
      Table 3Computed tomographic characteristics.
      ALCA (N ​= ​96)RCA (N ​= ​93)
      3 RCA could not be found.
      Coronary height, mm11.4 ​± ​5.213.8 ​± ​5.3
      VTC, mm
      VTC was assessed in type II and III.
      6.0 ​± ​2.35.2 ​± ​1.7
      VTC <4 ​mm, n (%)12 (16.2)11 (20.8)
      VTSTJ, mm
      VTSTJ was assessed in type IIIA and IIIC.
      5.6 ​± ​2.55.1 ​± ​1.7
      Narrow VTSTJ, n (%)8 (19.5)5 (21.7)
      VIVID Classification using the cut-off of <3.5 ​mm for VTSTJ, n (%)
       I21 (21.9)38 (40.9)
       IIA26 (27.1)23 (24.7)
       IIB4 (4.2)6 (6.5)
       IIIA32 (33.3)16 (17.2)
       IIIB8 (8.3)7 (7.5)
       IIIC5 (5.2)3 (3.2)
       Consider BASILICA, n (%)17 (17.7)16 (17.2)
      VIVID Classification using the cut-off of <2.5 ​mm for VTSTJ, n (%)
       I21 (21.9)38 (40.9)
       IIA26 (27.1)23 (24.7)
       IIB4 (4.2)6 (6.5)
       IIIA35 (36.5)19 (20.4)
       IIIB8 (8.3)7 (7.5)
       IIIC2 (2.1)0
       Consider BASILICA, n (%)14 (14.6)13 (14.0)
      VIVID Classification only in stented valve (N ​= ​84), n (%)
       I17 (20.2)32 (39.5)
       IIA26 (27.1)23 (24.7)
       IIB4 (4.2)6 (6.5)
       IIIA35 (36.5)19 (20.4)
       IIIB8 (8.3)7 (7.5)
       IIIC2 (2.1)0
       Consider BASILICA, n (%)14 (14.6)13 (14.0)
      LCA ​= ​left coronary artery; RCA ​= ​right coronary artery; VIVID = Valve-in-Valve International Data; VTC ​= ​virtual valve to coronary ostium; VTSTJ ​= ​virtual valve to sinotubular junction.
      a 3 RCA could not be found.
      b VTC was assessed in type II and III.
      c VTSTJ was assessed in type IIIA and IIIC.

      3.4 Procedural characteristics according to assigned VIVID classification

      Procedural characteristics in patients evaluated for the risk of coronary obstruction are presented in Table 2. Coronary protection was performed in 7 patients (7.3%) as they were assumed at high-risk of coronary obstruction due to a low offtake of the coronary arteries and/or a shallow sinus of Valsalva as assessed by aortography and preprocedural MDCT: 5 (5.2%) had BASILICA (4 single BASILICA for left coronary cusp and 1 double BASILICA for left and right coronary cusps), and 2 (2.1%) had wire protection. Three patients were formally however not considered at high risk of coronary obstruction according to the VIVID classification. The types of VIVID classification in the 7 cases are shown in Table 4. Of the 30 patients with high-risk anatomy, coronary protection using wire protection or BASILICA was performed in 3 patients. No coronary obstruction occurred in any of the patients with high-risk anatomy according to the VIVID classification.
      Table 4VIVID classification in patients undergoing coronary protection.
      Implanetd bioprosthetic valveReason for ViV-TAVITypes of coronary protectionVIVID type
      LCARCA
      Case 1Stented valve with externally mounted leafletsBioprosthetic regurgitationCoronary wire protection for LCAIIAI
      Case 2Sutureless valveBioprosthetic stenosisCoronary wire protection for LCAIIAIIB
      Case 3Stented valve with internally mounted leafletsBioprosthetic stenosisBASILICAIIIAIIIA
      Case 4Stented valve with externally mounted leafletsBioprosthetic stenosisBASILICAIIIAIIB
      Case 5Stented valve with internally mounted leafletsBioprosthetic stenosisBASILICAIIIAIIIA
      Case 6Stented valve with internally mounted leafletsBioprosthetic stenosisBASILICAIIIAIIIC
      Case 7Stented valve with internally mounted leafletsBioprosthetic stenosisBASILICAIIIBIIIA
      6 cases underwent coronary protection did not have adequate MDCT image for assessment of the VIVID classification.
      BASILICA ​= ​bioprosthetic or native aortic scallop intentional laceration to prevent iatrogenic coronary artery obstruction; LCA ​= ​left coronary artery; MDCT ​= ​multi-detector computed tomography; RCA ​= ​right coronary artery; ViV-TAVI ​= ​valve-in-valve transcatheter aortic valve implantation; VIVID = Valve-in-Valve International Data.

      4. Discussion

      The key findings of the present study can be summarized as follows. (1) In a prospective registry, coronary obstruction occurred in 2.9% of patients undergoing VIV-TAVI, and three in four obstructions occurred in patients with a stentless valve. (2) One-third of patients with a stented valve undergoing ViV-TAVI were deemed at high risk for coronary obstruction according to the VIVID classification. (3) Among patients with high-risk anatomy, coronary protection was performed in 10%. (4) Coronary obstruction occurred in none of the patients classified as high-risk patients according to the VIVID classification despite the absence of coronary protection.
      Coronary obstruction is a rare but life-threatening complication of TAVI.
      • Ribeiro H.B.
      • Rodes-Cabau J.
      • Blanke P.
      • et al.
      Incidence, predictors, and clinical outcomes of coronary obstruction following transcatheter aortic valve replacement for degenerative bioprosthetic surgical valves: insights from the VIVID registry.
      ,
      • Ribeiro H.B.
      • Webb J.G.
      • Makkar R.R.
      • et al.
      Predictive factors, management, and clinical outcomes of coronary obstruction following transcatheter aortic valve implantation: insights from a large multicenter registry.
      The incidence of coronary obstruction following ViV-TAVI has been reported as 2.0–3.5%,
      • Dvir D.
      • Webb J.G.
      • Bleiziffer S.
      • et al.
      Transcatheter aortic valve implantation in failed bioprosthetic surgical valves.
      ,
      • Ribeiro H.B.
      • Rodes-Cabau J.
      • Blanke P.
      • et al.
      Incidence, predictors, and clinical outcomes of coronary obstruction following transcatheter aortic valve replacement for degenerative bioprosthetic surgical valves: insights from the VIVID registry.
      ,
      • Dvir D.
      • Webb J.
      • Brecker S.
      • et al.
      Transcatheter aortic valve replacement for degenerative bioprosthetic surgical valves: results from the global valve-in-valve registry.
      which is considerably higher than rates of coronary obstruction in TAVI for native aortic valve stenosis (<1%).
      • Ribeiro H.B.
      • Webb J.G.
      • Makkar R.R.
      • et al.
      Predictive factors, management, and clinical outcomes of coronary obstruction following transcatheter aortic valve implantation: insights from a large multicenter registry.
      In line with previous studies, the incidence of coronary obstruction in the present study was 2.9%, and 3 out of 4 coronary obstructions occurred following ViV-TAVI in a stentless valve.
      • Ribeiro H.B.
      • Rodes-Cabau J.
      • Blanke P.
      • et al.
      Incidence, predictors, and clinical outcomes of coronary obstruction following transcatheter aortic valve replacement for degenerative bioprosthetic surgical valves: insights from the VIVID registry.
      ,
      • Duncan A.
      • Moat N.
      • Simonato M.
      • et al.
      Outcomes following transcatheter aortic valve replacement for degenerative stentless versus stented bioprostheses.
      There are different mechanisms for coronary obstruction in ViV-TAVI procedures. In patients with low coronary offtake and a tubular shape of the sinus of Valsalva, coronary ostia can be directly obstructed by the leaflets of the bioprosthesis. Alternatively, coronary ostia can be obstructed indirectly by sinus sequestration in patients with a narrow and low sinotubular junction.
      Dedicated risk assessment tools and methods for prevention have been proposed. A short VTC distance (<3–4 ​mm) has been suggested as an important predictor of coronary obstruction following ViV-TAVI based on data from the VIVID registry.
      • Ribeiro H.B.
      • Rodes-Cabau J.
      • Blanke P.
      • et al.
      Incidence, predictors, and clinical outcomes of coronary obstruction following transcatheter aortic valve replacement for degenerative bioprosthetic surgical valves: insights from the VIVID registry.
      Moreover, recent CT-based simulation analyses applied hypothesis-based criteria using a VTSTJ distance for the assessment of the risk of coronary obstruction due to sinus sequestration.
      • Ochiai T.
      • Chakravarty T.
      • Yoon S.H.
      • et al.
      Coronary access after TAVR.
      BASILICA is a promising, although technically demanding, technique to mitigate the risk of coronary obstruction among patients undergoing ViV-TAVI.
      • Khan J.M.
      • Rogers T.
      • Greenbaum A.B.
      • et al.
      Transcatheter electrosurgery: JACC state-of-the-art review.
      In a feasibility study and international registry, BASILICA was successful in 86.9–93.3%, and rates of coronary obstruction, death, and disabling stroke at 30 days were 0–4.7%, 2.8–3.3% and 0.5–3.3%, respectively.
      • Khan J.M.
      • Greenbaum A.B.
      • Babaliaros V.C.
      • et al.
      The BASILICA trial: prospective multicenter investigation of intentional leaflet laceration to prevent TAVR coronary obstruction.
      ,
      • Khan J.M.
      • Babaliaros V.C.
      • Greenbaum A.B.
      • et al.
      Preventing coronary obstruction during transcatheter aortic valve replacement.
      ,
      • Khan J.M.
      • Greenbaum A.B.
      • Babaliaros V.C.
      • et al.
      BASILICA trial: one-year outcomes of transcatheter electrosurgical leaflet laceration to prevent TAVR coronary obstruction.
      The VIVID classification has been proposed to systematically assess the risk of coronary obstruction in ViV-TAVI for stented valves and to evaluate the need for BASILICA based on preprocedural MDCT.
      • Tang G.H.L.
      • Komatsu I.
      • Tzemach L.
      • et al.
      Risk of coronary obstruction and the need to perform BASILICA: the VIVID classification.
      The present study is the first to evaluate the validity of the VIVID classification in a prospective registry and showed that coronary obstruction was rare even in patients with high-risk anatomy (VIVID type IIB, IIIB, or IIIC) and in the absence of coronary protection. The VIVID classification is a useful instrument to sensitize operators for coronary obstruction in a particular subset of patients planned for ViV-TAVI, and identifies anatomic constellations in which coronary protection may be warranted. While the classification provides a solid basis for the discussion of different treatment strategies, a tailored approach continues to be key, and the decision to perform coronary protection needs to be individualized. In addition, current algorithm does not take account for the procedural factor, including implantation depth and THV tilting. Multimodality imaging and 3D printing may be used to refine this process. Importantly the VIVID-classification is not applicable to patients with stentless valves who have the greatest risk of coronary obstruction with ViV-TAVI.

      4.1 Study limitations

      The findings of our study should be interpreted in light of several limitations. First, this is a single-center study, and the number of cases was relatively small. In turn, to our knowledge, our study is the first to validate the proposed algorithm in a dedicated Corelab. Second, although the risk of coronary obstruction was Core laboratory evaluated, more than 10% of patients were excluded from the CT analysis due to inadequate MDCT images. Furthermore, we did not routinely perform post-TAVI CT which would have allowed us to confirm the VTC and VTSTJ. Third, the results of the present study reflect the experience of a single high-volume center and may not be generalizable to other heart centers. Finally, the present study included only patients who underwent ViV-TAVI, resulting in selection bias. Due to its retrospective nature, the number of patients who were turned down for TAVI and underwent surgical aortic valve replacement or conservative management due to the risk of coronary obstruction were not available. Thus, the risk of coronary obstruction was likely underestimated.

      5. Conclusion

      Three in four coronary obstructions following ViV-TAVI occurred in patients with a stentless valve while none of the patients underogoing ViV-TAVI for stented bioprostheses at high risk according to the VIVID classification had coronary obstruction. Refined tools are required to assess the risk of coronary obstruction, particularly for patients with stentless valves.

      Funding

      None.

      Declaration of competing interest

      Dr. Windecker reports research, travel or educational grants to the institution from Abbott, Abiomed, Amgen, Astra Zeneca, Bayer, Biotronik, Boehringer Ingelheim, Boston Scientific, Bristol Myers Squibb, Cardinal Health, CardioValve, Corflow Therapeutics, CSL Behring, Daiichi Sankyo, Edwards Lifesciences, Guerbet, InfraRedx, Janssen-Cilag, Johnson & Johnson, Medicure, Medtronic, Merck Sharp & Dohm, Miracor Medical, Novartis, Novo Nordisk, Organon, OrPha Suisse, Pfizer, Polares, Regeneron, Sanofi-Aventis, Servier, Sinomed, Terumo, Vifor, V-Wave.
      Dr. Windecker serves as unpaid advisory board member and/or unpaid member of the steering/executive group of trials funded by Abbott, Abiomed, Amgen, Astra Zeneca, Bayer, Boston Scientific, Biotronik, Bristol Myers Squibb, Edwards Lifesciences, Janssen, MedAlliance, Medtronic, Novartis, Polares, Recardio, Sinomed, Terumo, V-Wave and Xeltis, but has not received personal payments by pharmaceutical companies or device manufacturers. He is also member of the steering/executive committee group of several investigator-initiated trials that receive funding by industry without impact on his personal remuneration. Dr. Pilgrim reports research grants to the institution from Edwards Lifesciences, Boston Scientifc and Biotronik, personal fees from Biotronik, Boston Scientific, Metronic, Abbott, and HighLife SAS. Dr. Reineke reports travel expenses from Abbott, Edwards Lifesciences and Medtronic. Dr. Stortecky reports research grants to the institution from Edwards Lifesciences, Medtronic, Boston Scientific and Abbott and personal fees from Boston Scientific, Teleflex, and BTG. Dr. Okuno reports speaker fees from Abbott. All other authors have no relationships relevant to the contents of this article to disclose.

      References

        • Dvir D.
        • Webb J.G.
        • Bleiziffer S.
        • et al.
        Transcatheter aortic valve implantation in failed bioprosthetic surgical valves.
        JAMA. 2014; 312: 162-170
        • Tuzcu E.M.
        • Kapadia S.R.
        • Vemulapalli S.
        • et al.
        Transcatheter aortic valve replacement of failed surgically implanted bioprostheses: the STS/ACC registry.
        J Am Coll Cardiol. 2018; 72: 370-382
        • Webb J.G.
        • Mack M.J.
        • White J.M.
        • et al.
        Transcatheter aortic valve implantation within degenerated aortic surgical bioprostheses: PARTNER 2 valve-in-valve registry.
        J Am Coll Cardiol. 2017; 69: 2253-2262
        • Webb J.G.
        • Murdoch D.J.
        • Alu M.C.
        • et al.
        3-Year outcomes after valve-in-valve transcatheter aortic valve replacement for degenerated bioprostheses: the PARTNER 2 registry.
        J Am Coll Cardiol. 2019; 73: 2647-2655
        • Deeb G.M.
        • Chetcuti S.J.
        • Reardon M.J.
        • et al.
        1-Year results in patients undergoing transcatheter aortic valve replacement with failed surgical bioprostheses.
        JACC Cardiovasc Interv. 2017; 10: 1034-1044
        • Bleiziffer S.
        • Simonato M.
        • Webb J.G.
        • et al.
        Long-term outcomes after transcatheter aortic valve implantation in failed bioprosthetic valves.
        Eur Heart J. 2020; 41: 2731-2742
        • Hahn R.T.
        • Webb J.
        • Pibarot P.
        • et al.
        5-Year follow-up from the PARTNER 2 aortic valve-in-valve registry for degenerated aortic surgical bioprostheses.
        JACC Cardiovasc Interv. 2022; 15: 698-708
        • Carroll J.D.
        • Mack M.J.
        • Vemulapalli S.
        • et al.
        STS-ACC TVT registry of transcatheter aortic valve replacement.
        J Am Coll Cardiol. 2020; 76: 2492-2516
        • Otto C.M.
        • Nishimura R.A.
        • Bonow R.O.
        • et al.
        ACC/AHA guideline for the management of patients with valvular heart disease: a report of the American college of cardiology/American heart association joint committee on clinical practice guidelines.
        J Am Coll Cardiol. 2020; 77 (2021): e25-e197
        • Vahanian A.
        • Beyersdorf F.
        • Praz F.
        • et al.
        2021 ESC/EACTS Guidelines for the management of valvular heart disease.
        Eur Heart J. 2022; 43: 561-632
        • Dvir D.
        • Leipsic J.
        • Blanke P.
        • et al.
        Coronary obstruction in transcatheter aortic valve-in-valve implantation: preprocedural evaluation, device selection, protection, and treatment.
        Circ Cardiovasc Interv. 2015; 8e002079
        • Okuno T.
        Risk of "future" coronary obstruction: a key factor in patient-tailored lifetime management of aortic stenosis.
        JACC Cardiovasc Interv. 2022; 15: 725-727
        • Ribeiro H.B.
        • Rodes-Cabau J.
        • Blanke P.
        • et al.
        Incidence, predictors, and clinical outcomes of coronary obstruction following transcatheter aortic valve replacement for degenerative bioprosthetic surgical valves: insights from the VIVID registry.
        Eur Heart J. 2018; 39: 687-695
        • Khan J.M.
        • Greenbaum A.B.
        • Babaliaros V.C.
        • et al.
        The BASILICA trial: prospective multicenter investigation of intentional leaflet laceration to prevent TAVR coronary obstruction.
        JACC Cardiovasc Interv. 2019; 12: 1240-1252
        • Mercanti F.
        • Rosseel L.
        • Neylon A.
        • et al.
        Chimney stenting for coronary occlusion during TAVR: insights from the chimney registry.
        JACC Cardiovasc Interv. 2020; 13: 751-761
        • Tang G.H.L.
        • Komatsu I.
        • Tzemach L.
        • et al.
        Risk of coronary obstruction and the need to perform BASILICA: the VIVID classification.
        EuroIntervention. 2020; 16: e757-e759
        • Stortecky S.
        • Franzone A.
        • Heg D.
        • et al.
        Temporal trends in adoption and outcomes of transcatheter aortic valve implantation: a SwissTAVI Registry analysis.
        Eur Heart J Qual Care Clin Outcomes. 2019; 5: 242-251
        • Okuno T.
        • Brugger N.
        • Asami M.
        • et al.
        Clinical impact of mitral calcium volume in patients undergoing transcatheter aortic valve implantation.
        J Cardiovasc Comput Tomogr. 2021; 15: 356-365
        • Blanke P.
        • Weir-McCall J.R.
        • Achenbach S.
        • et al.
        Computed tomography imaging in the context of transcatheter aortic valve implantation (TAVI)/Transcatheter aortic valve replacement (TAVR): an expert consensus document of the society of cardiovascular computed tomography.
        J Cardiovasc Comput Tomogr. 2019; 13: 1-20
        • Tarantini G.
        • Dvir D.
        • Tang G.H.L.
        Transcatheter aortic valve implantation in degenerated surgical aortic valves.
        EuroIntervention. 2021; 17: 709-719
        • Leon M.B.
        • Piazza N.
        • Nikolsky E.
        • et al.
        Standardized endpoint definitions for transcatheter aortic valve implantation clinical trials: a consensus report from the valve academic research consortium.
        J Am Coll Cardiol. 2011; 57: 253-269
        • Kappetein A.P.
        • Head S.J.
        • Généreux P.
        • et al.
        Updated standardized endpoint definitions for transcatheter aortic valve implantation: the Valve Academic Research Consortium-2 consensus document.
        J Am Coll Cardiol. 2012; 60: 1438-1454
        • Ribeiro H.B.
        • Webb J.G.
        • Makkar R.R.
        • et al.
        Predictive factors, management, and clinical outcomes of coronary obstruction following transcatheter aortic valve implantation: insights from a large multicenter registry.
        J Am Coll Cardiol. 2013; 62: 1552-1562
        • Dvir D.
        • Webb J.
        • Brecker S.
        • et al.
        Transcatheter aortic valve replacement for degenerative bioprosthetic surgical valves: results from the global valve-in-valve registry.
        Circulation. 2012; 126: 2335-2344
        • Duncan A.
        • Moat N.
        • Simonato M.
        • et al.
        Outcomes following transcatheter aortic valve replacement for degenerative stentless versus stented bioprostheses.
        JACC Cardiovasc Interv. 2019; 12: 1256-1263
        • Ochiai T.
        • Chakravarty T.
        • Yoon S.H.
        • et al.
        Coronary access after TAVR.
        JACC Cardiovasc Interv. 2020; 13: 693-705
        • Khan J.M.
        • Rogers T.
        • Greenbaum A.B.
        • et al.
        Transcatheter electrosurgery: JACC state-of-the-art review.
        J Am Coll Cardiol. 2020; 75: 1455-1470
        • Khan J.M.
        • Babaliaros V.C.
        • Greenbaum A.B.
        • et al.
        Preventing coronary obstruction during transcatheter aortic valve replacement.
        JACC Cardiovasc Interv. 2021; 14: 941-948
        • Khan J.M.
        • Greenbaum A.B.
        • Babaliaros V.C.
        • et al.
        BASILICA trial: one-year outcomes of transcatheter electrosurgical leaflet laceration to prevent TAVR coronary obstruction.
        Circ Cardiovasc Interv. 2021; 14e010238